Contact us | Sign in | U.S.A. U.S.A. (Change) - English

U19 Socketed IC

Urgent: Medical Device Recall Notification

Adobe PDF U19 Socketed IC
PDF (79 Kb)
Adobe PDF U9 Socketed IC and U19 Socketed IC
PDF (98 Kb)
Adobe PDF Grey IUI Connectors, U9 Socketed IC, U19 Socketed IC
PDF (121 Kb)

Related information

Adobe PDF U19 Related Serial Numbers
PDF (523 Kb)

Contact information:

Cardinal Health Recall Center
3750 Torrey View Court
San Diego, CA 92130
800.625.6627
  • Affected Units: Alaris® Pump modules manufactured prior to September 8, 2005
  • Issue: The U19 Socketed IC is located on the Alaris® Pump module logic board and is responsible for monitoring functions. Over time the U19 Socketed IC fit may become loose due to factors such as vibration.
  • Recall Information and Q&A: For the most current information or answers to questions, please contact the Cardinal Health Recall Center at 800.625.6627.
  • Adverse Reports: If you have any adverse reports related to the Alaris® System, please use our main business line to contact Cardinal Health Customer Advocacy at 800.854.7128, Option 1, Option 1, then Option 3, or via email at customerfeedback@cardinalhealth.com.